Agilent Technologies 81480A Stud Sensor User Manual


 
Test Record Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 105
Agilent 81480A Performance Test
Page 7 of 13
Model Agilent 81480A Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty: ±0.10 dB
Output 1 Output 2
Wavelength Setting Power Measured Minimum Specification Power Measured Minimum Specification
1370.000 nm dBm 13.00 dBm dBm 3.00 dBm
1380.000 nm dBm 13.00 dBm dBm 3.00 dBm
1390.000 nm dBm 13.00 dBm dBm 3.00 dBm
1400.000 nm dBm 13.00 dBm dBm 3.00 dBm
1410.000 nm dBm 13.00 dBm dBm 3.00 dBm
1420.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1430.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1440.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1450.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1460.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1470.000 nm dBm 7.00 dBm dBm + 3.00 dBm
1480.000 nm dBm 13.00 dBm dBm 3.00 dBm