Agilent Technologies 81480A Stud Sensor User Manual


 
Performance Tests Test Record
108 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
Agilent 81480A Performance Test
Page 10 of 13
Model Agilent 81480A Tunable Laser Report No. ________ Date_______
Power Flatness
Wavelength
Low SSE Output 1 High Power Output 2
P = 13 dBm
P = 3dBm
ATT = 0 dB
P = 3dBm
ATT = 57 dB
Power Deviation Power Deviation Power Deviation
Start = REF 1420.000 nm 0.00 dB 0.00 dB 0.00 dB
1425.000 nm dB dB dB
1430.000 nm dB dB dB
1435.000 nm dB dB dB
1440.000 nm dB dB dB
1445.000 nm dB dB dB
1450.000 nm dB dB dB
1455.000 nm dB dB dB
1460.000 nm dB dB dB
1465.000 nm dB dB dB
1470.000 nm dB dB dB
1480.000 nm dB dB dB
Maximum deviation dBdBdB
Minimum deviation dB dB dB
Flatness = MaximumMinimum Deviation dB dB dB
Specification 0.40 dBpp 0.60 dBpp 0.60 dBpp
Measurement Uncertainty ± 0.1 dB ± 0.1 dB ± 0.1 dB