Agilent Technologies 81480A Stud Sensor User Manual


 
Test Record Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 119
Agilent 81680A Performance Test
Page 7 of 13
Model Agilent 81680A Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty: ±0.10 dB
Output 1 Output 2
Wavelength Setting Power Measured Minimum Specification Power Measured Minimum Specification
1460.000 nm dBm 13.00 dBm dBm 3.00 dBm
1470.000 nm dBm 13.00 dBm dBm 3.00 dBm
1480.000 nm dBm 10.00 dBm dBm + 1.00 dBm
1490.000 nm dBm 10.00 dBm dBm + 1.00 dBm
1500.000 nm dBm 10.00 dBm dBm + 1.00 dBm
1510.000 nm dBm 10.00 dBm dBm + 1.00 dBm
1520.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1530.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1540.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1550.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1560.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1570.000 nm dBm 6.00 dBm dBm + 5.00 dBm
1580.000 nm dBm 10.00 dBm dBm + 1.00 dBm