Agilent Technologies 81480A Stud Sensor User Manual


 
Performance Tests Test Record
160 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
Agilent 81642A Performance Test
Page 8 of 12
Model Agilent 81642A Tunable Laser Report No. ________ Date_______
Power Linearity 81642A #003 Upper Power Levels
Maximum Power Linearity at current setting ______dB
Minimum Power Linearity at current setting ______dB
Total Power Linearity = (Max Power Linearity Min Power Linearity) ______dBpp
Specification 1.0 dBpp
Typical 0.6 dBpp
Measurement Uncertainty ±0.05 dB
Power Setting
from start
Measured Relative Pow-
er from start
Power reduction
from start
Power Linearity at cur-
rent setting
Start = REF + 4.5 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 3.5 dBm dB + 1.00 dB = dB
+ 2.5 dBm dB + 2.00 dB = dB
+ 1.5 dBm dB + 3.00 dB = dB
+ 0.5 dBm dB + 4.00 dB = dB
0.5 dBm dB + 5.00 dB = dB
1.5 dBm dB + 6.00 dB = dB
2.5 dBm dB + 7.00 dB = dB
3.5 dBm dB + 8.00 dB = dB
4.5 dBm dB + 9.00 dB = dB