Te st Re c or d
Results of the performance test may be tabulated in the Test Record
provided at the end of the test procedures. It is recommended that you
fill out the Test Record and refer to it while doing the test. Since the
test limits and setup information are printed on the Test Record for
easy reference, the record can also be used as an abbreviated test
procedure (if you are already familiar with the test procedures). The
Test Record can also be used as a permanent record and may be
reproduced without written permission from Agilent Technologies.
Test Failure
Always ensure that you use the correct cables and adapters, and that
all connectors are undamaged and extremely clean.
If the Agilent 81480A or Agilent 81680A/40A/82A/42A/89A Tunable
Laser module fails any performance test, return the instrument to the
nearest Agilent Technologies Sales/Service Office for repair.
Instrument Specification
Specifications are the performance characteristics of the instrument
which are certified. These specifications, listed in “Specifications” on
page 39, are the performance standards or limits against which the
Agilent 81480A and Agilent 81680A/40A/82A/42A/89A Tunable Laser
modules can be tested.
The specifications also list some supplemental characteristics of the
Agilent 81480A and Agilent 81680A/40A/82A/42A/89A Tunable Laser
modules. Supplemental characteristics should be considered as
additional information.
Any changes in the specifications due to manufacturing changes,
design, or traceability to the National Institute of Standards and
Technology (NIST), will be covered in a manual change supplement, or
revised manual. Such specifications supersede any that were
previously published.