Agilent Technologies 81640A Stud Sensor User Manual


 
Performance Tests Performance Test Instructions
70 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
6 Set the wavelength and power for each Tunable Laser module to the
values given in Table Table 10.
7 Press the key beside the laser output to switch on the laser output.
NOTE The laser output is limited to its maximum possible value at this
wavelength, the display will probably show ExP.
8 Set the wavelength of the HP 8153A to the same as your Tunable
Laser module, as given in Table Table 10.
9 Measure the output power with the HP 8153A and note the result
for this wavelength in the test record.
10 Increase the λ, output wavelength, of the Tunable Laser module to
the next value given in the test record.
11 Increase the wavelength of the HP 8153A to the same value.
12 Note the measured power in the test record for each wavelength
13 Repeat item 10 to item 12 for the full wavelength range
14 If you are using the Agilent 81680A, Agilent 81480A, or the
Agilent 81640A Tunable Laser module:
Connect the output fiber to Output 2, the High Power output,
remember to calibrate the Agilent 81001FF Attenuation Filter and
set <Optical Output> to <High Power (2)>.
Then, perform steps 6 through 12 for the full wavelength range.
Table 10 Reference Wavelength and Power Values for Maximum Output Power Tests
Module Wavelength [λ] Power [P]
Agilent 81480A - Output 1 1370.200 nm +0.00 dBm
Agilent 81480A - Output 2 1370.200 nm +10.00 dBm
Agilent 81680A - Output 1 1460.000 nm +0.00 dBm
Agilent 81680A - Output 2 1460.000 nm +10.00 dBm
Agilent 81640A - Output 1 1510.000 nm +0.00 dBm
Agilent 81640A - Output 2 1510.000 nm +10.00 dBm
Agilent 81682A 1460.000 nm +10.00 dBm
Agilent 81642A 1510.000 nm +10.00 dBm
Agilent 81689A 1525.000 nm +10.00 dBm