Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 85
Figure 25 Test Setup for Source Spontaneous Emission Test
2 Determine the filter transmission characteristics:
a Check center wavelength, l
FBG
, of the Fiber Bragg Grating. This
wavelength is printed on its label, for example, 1520.5 nm. This
value relates to measurements performed in a vacuum.
b Set the Optical Spectrum Analyzer:
Set the Span to 8 nm. Press Span and enter the value.
Set the center wavelength to λ
FBG
0.5 nm. Press Center and enter
the value.
Set the reference level to 0 dBm. Press [AMPL], press [Ref LVL],
and enter the value.
Set the Sensitivity to -68 dBm. Press [AMPL], press
[SENS AUTO MAN
], and enter the value.
Set the resolution bandwidth to 0.1 nm. Press [BW Swp], and enter
the value.
c Set the Tunable Laser module
Set [λ], the wavelength, to λ
FBG
1 nm, for example,
1520.5 nm 1 nm = 1519.5 nm.
Set [P], the output power, to the value in Table Table 18.
d Press the key beside the laser output to switch on the laser
output.
e Check and note the peak power level displayed by the OSA and
the wavelength at the peak power. Press Peak Search in the
Marker field.
Agilent 8164A Lightwave
Measurement System
Tunable Laser
For #021, #071: use 81000FI and adapter DIN/DIN 1005-0255 and 81113PC and 81000FI
For #022, #072: use 81000SI and 81000FI
Optical Spectrum Analyzer
Fiber
Grating
Bragg
Table 18 Output Power Setting - Low SSE Output
Tunable Laser Module
Power [P]
Agilent 81480A - Output 1 -7
dBm
Agilent 81680A - Output 1 -6 dBm
Agilent 81640A - Output 1 -7 dBm