Agilent Technologies 81640A Stud Sensor User Manual


 
Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 93
Sum of all SSE power levels:
in lower transmission band ________ pW (1)
in upper transmission band ________ pW (2)
Sum of all SSE power levels in transmission bands, add results in (1)
and (2)
power_trans = ___________ pW
11 Determine SSE power inside the attenuation band by interpolation:
a Check the power measured at OSA_l2 and OSA_l3.
b Mark that power value which is the largest of both and note it as
power_ OSA_l2,3_max
c Calculate the power inside the attenuation band by using
power_att = 1/2 ¥ power_ OSA_l2,3_max
= ___________ 10
-12
W = ___________ pW
NOTE Note all the power values in [pW], where 1 pW = 10
12
W.
+3 nm pW + 3 nm pW
+4 nm pW + 4 nm pW
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
+11 nm pW +11 nm pW
+12 nm pW +12 nm pW
+13 nm pW +13 nm pW
+14 nm
(relates to OSA_λ2)
pW +14 nm
(relates to OSA_λ4)
pW
Lower transmission band
OSA_λ1 to OSA_λ2
Upper transmission band
OSA_λ3 to OSA_λ4
Relative Wavelength,
Increments
from λ_1
SSE_power mea-
sured
Relative Wavelength,
Increments
from λ_3
SSE_power mea-
sured