Agilent Technologies 81642A Stud Sensor User Manual


 
Test Record Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 163
Agilent 81642A Performance Test
Page 11 of 12
Model Agilent 81642A Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - 81642A
Measurement Uncertainty: ±0.20 dB
Agilent 81642A
Standard
without #003
Agilent 81642A
Option #003
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Specification 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ± 0.005 dB ± 0.005 dB
Agilent 81642A Standard
without #003
Agilent 81642A
Option #003
Wavelength Output
Power
Results Maximum
Specification
Output
Power
Results Maximum
Specification
1510 nm 3.00 dBm dB 35 dB 4.50 dBm dB 35 dB
1520 nm +2.00 dBm dB 40 dB +0.50 dBm dB 40 dB
1530 nm +4.00 dBm dB 45 dB +2.50 dBm dB 45 dB
1540 nm +4.00 dBm dB 45 dB +2.50 dBm dB 45 dB
1550 nm +4.00 dBm dB 45 dB +2.50 dBm dB 45 dB
1560 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1570 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1580 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1590 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1600 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1610 nm +6.00 dBm dB 45 dB +4.50 dBm dB 45 dB
1620 nm +2.00 dBm dB 40 dB +0.50 dBm dB 40 dB
1630 nm
3.00 dBm dB 35 dB 4.50 dBm dB 35 dB
1640 nm
3.00 dBm dB 30 dB 4.50 dBm dB 30 dB