Performance Tests Performance Test Instructions
68 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
5 Check if the display of HP 8153A reads 0 dBm ±2 dBm.
6 Select dB as the power units of the HP 8153A.
7 Press Disp>Ref on the HP 8153A.
8 Attach the Agilent 81001FF Attenuation Filter to the Optical Head
as shown in Figure 19. Move the patchcord as little as possible,
keeping the laser activated.
Figure 19 Test Setup for Calibrating the Agilent 81001FF Attenuation Filter
9 On the HP 8153A Lightwave Multimeter, move to the power
measurement channel where the HP 81533B is installed. Press
Parameter until CAL is displayed. Using the Modify keys, change
the CAL value until the display shows 0.00 dB.
10 For further measurements using this Attenuation Filter, you have to
set the calibration factor of the HP 8153A to this value to get the
absolute power reading.
Agilent 8164A Lightwave
Measurement System
HP 81524A
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted
Connector
Optical Head
Agilent 81001FF81000SA
HP 81533B
HP 8153A Lightwave Multimeter