Agilent Technologies 81642A Stud Sensor User Manual


 
Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 71
Power Linearity
Power Linearity - Low Power Test
To measure the power linearity of the Low SSE output, Output 1, of
the Agilent 81480A, Agilent 81680A, or the of the Agilent 81640A:
1 Set up the equipment as shown in Figure 21.
Figure 21 Test Setup for Low Power Linearity Tests
2 Move to the Tunable Laser channel of the Agilent 8164A Lightwave
Measurement System and press [Menu].
3 Set the menu parameters to the values shown in Table Table 5.
<Power Mode> does not apply.
4 Set the wavelength and power for each Tunable Laser module to the
values given in Table Table 11.
5 If you are using the Agilent 81680A, Agilent 81480A, or the
Agilent 81640A Tunable Laser module:
Connect the output fiber to Output 1, the Low SSE output.
Set <Optical Output> to <Low SSE (1)>.
6 Make sure the optical output is switched off.
Agilent 8164A Lightwave
Measurement System
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted
Connector
HP 81532A
HP 8153A Lightwave Multimeter
Power Sensor
Table 11 Wavelength and Power Settings for Low Power Linearity Tests
Module Wavelength [λ] Power [P]
Agilent 81480A - Output 1 1430.000 nm 7.00 dBm
Agilent 81680A - Output 1 1540.000 nm 6.00 dBm
Agilent 81640A - Output 1 1560.000 nm 7.00 dBm