Agilent Technologies 81642A Stud Sensor User Manual


 
Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 75
Power Linearity - Test Using Attenuation
Follow the steps below to measure the power linearity (while using
attenuation) of any one of the following:
Output 2, the High Power output, of the Agilent 81480A
Output 2, the High Power output, of the Agilent 81680A
Output 2, the High Power output, of the Agilent 81640A
Agilent 81682A #003
Agilent 81642A #003
1 Set up the equipment as shown in Figure 21.
2 Move to the Tunable Laser channel of the Agilent 8164A Lightwave
Measurement System and press [Menu].
3 Set the menu parameters to the values shown in Table Table 5.
For Agilent 81480A, Agilent 81680A, Agilent 81640A,
Agilent 81642A#003 Agilent 81682A#003:
Set <Power Mode> to <Manual Att>.
4 Set the wavelength and power for each Tunable Laser module to the
values given in Table Table 13.
NOTE If you use the Agilent 81480A Output 2, Agilent 81680A Output 2, or
Agilent 81640A Output 2 with attenuation, use the table “Power
Linearity Output 2, High Power by attenuator” on page 121 or “Power
Linearity Output 2, High Power by Attenuator” on page 135
respectively.
If you use the Agilent 81682A #003 with attenuation, use the table
“Power Linearity 81682A #003 by Attenuator” on page 149.
If you use the Agilent 81642A #003 with attenuation, use the table
“Power Linearity 81642A #003 by Attenuator” on page 161.
5 Perform the steps 5 to 16 of the “Power Linearity - Low Power Test”
on page 71.
Table 13 Wavelength and Power Settings for High Power Linearity Tests with
Attenuation
Module Wavelength [λ] Power [P] Attenuation [Atten]
Agilent 81480A - Output 2 1430.000 nm +0.000 dBm 0.000 dB
Agilent 81680A - Output 2 1540.000 nm +0.000 dBm 0.000 dB
Agilent 81640A - Output 2 1560.000 nm +0.000 dBm 0.000 dB
Agilent 81682A #003 1540.000 nm +0.000 dBm 0.000 dB
Agilent 81642A #003 1580.000 nm +0.000 dBm 0.000 dB