Test Record Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 147
Agilent 81682A Performance Test Page 7 of 12
Model Agilent 81682A Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty: ± 0.10 dB
Power Linearity - 81682A
Maximum Power Linearity at current setting ______dB
Minimum Power Linearity at current setting ______dB
Total Power Linearity = (Max Power Linearity Min Power Linearity) ______dBpp
Specification 0.5 dBpp
typical 0.2 dBpp
Measurement Uncertainty ±0.05 dB
Agilent 81682A Agilent 81682A #003
Wavelength Setting Power Measured Minimum Specification Power Measured Minimum Specification
1460.000 nm dBm 3.00 dBm dBm 4.50 dBm
1470.000 nm dBm 3.00 dBm dBm 4.50 dBm
1480.000 nm dBm + 2.00 dBm dBm + 0.50 dBm
1490.000 nm dBm + 2.00 dBm dBm + 0.50 dBm
1500.000 nm dBm + 2.00 dBm dBm + 0.50 dBm
1510.000 nm dBm + 2.00 dBm dBm + 0.50 dBm
1520.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1530.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1540.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1550.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1560.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1570.000 nm dBm + 6.00 dBm dBm + 4.50 dBm
1580.000 nm dBm + 2.00 dBm dBm + 0.50 dBm
Power Setting from
start
Measured Relative Power
from start
Power reduction from
start
Power Linearity at current
setting
Start = REF + 6.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 5.0 dBm dB + 1.00 dB = dB
+ 4.0 dBm dB + 2.00 dB = dB
+ 3.0 dBm dB + 3.00 dB = dB
+ 2.0 dBm dB + 4.00 dB = dB
+ 1.0 dBm dB + 5.00 dB = dB
+ 0.0 dBm dB + 6.00 dB = dB
1.0 dBm dB + 7.00 dB = dB
2.0 dBm dB + 8.00 dB = dB
3.0 dBm dB + 9.00 dB = dB