Agilent Technologies 81689A Stud Sensor User Manual


 
Performance Tests Test Record
168 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
Agilent 81689A Performance Test Page 4 of 6
Model Agilent 81689A Tunable Laser Report No. ________ Date_______
Absolute Wavelength Accuracy
Result
Largest Value of Deviation (= largest value of either Largest Maximum
Deviation or Smallest Minimum Deviation)
Absolute Wavelength Accuracy________nm
Specification 1.0 nm
0.6 nm typical
Measurement Uncertainty: ±0.6 pm
Wavelength Repeatability
Measurement Uncertainty: ±0.1 pm
Repeatability of
1525.000 nm (= reference) Measurement Result
Repeatability of
1550.000 nm (= reference) Measurement Result
Initial Setting REF = nm Initial Setting REF = nm
from 1535.000 nm to REF nm from 1525.000 nm to REF nm
from 1540.000 nm to REF nm from 1535.000 nm to REF nm
from 1550.000 nm to REF nm from 1540.000 nm to REF nm
from 1560.000 nm to REF nm from 1560.000 nm to REF nm
from 1575.000 nm to REF nm from 1575.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured wavelength nm smallest measured wavelength nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured wavelength = largest measured wavelength - smallest measured wavelength
Specification 0.10 nm Specification 0.10 nm
Repeatability of
1575.000 nm (= reference)
Measurement
Result
Initial Setting REF = nm
from 1525.000 nm to REF nm
from 1535.000 nm to REF nm
from 1540.000 nm to REF nm
from 1550.000 nm to REF nm
from 1560.000 nm to REF nm
largest measured wavelength nm
smallest measured wavelength nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured wavelength
Specification 0.10 nm