Agilent Technologies 81689A Stud Sensor User Manual


 
Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 61
Performance Test Instructions
NOTE Make sure that all fiber connectors are clean.
Turn the instruments on, enable the laser and allow the
instruments to warm up.
Ensure that the Device Under Test (DUT) and all the test
equipment is held within the environmental specifications given
in “Specifications” on page 39
General Test Setup
Insert your Tunable Laser module into the Agilent 8164A Lightwave
Measurement System. Insert an Agilent 81480A, Agilent 81680A,
Agilent 81640A, Agilent 81682A, or Agilent 81642A Tunable Laser
module from the rear. Insert an Agilent 81689A Tunable Laser module
from the front into slot 1 of the Agilent 8164A Lightwave Measurement
System.
Wavelength Tests
NOTE When performing wavelength tests, zero the Tunable Laser first.
Move to Channel 0, press [Menu], select <λ Zeroing>.
Zeroing takes approximately 2 minutes.
Connect the Tunable Laser module to the Wavelength Meter as shown
in Figure 17.
If you use the Agilent 81480A, Agilent 81680A or Agilent 81640A
Tunable Laser module, connect the Output 2, the high power output.
Figure 17 Test Setup for Wavelength Tests
Agilent 8164A Lightwave
Measurement System
Wavelength Meter
Tunable Laser
For #021, #071: use 81000AI and 81101PC
For #022, #072: use 81000SI and 81113PC
Straight
Connector