Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 67
Power Tests
Calibration of the Agilent 81001FF Attenuation Filter
NOTE When an HP 81524A Optical Head is used in conjunction with a
Agilent 81001FF Attenuation Filter, it is absolutely necessary that you
calibrate the Agilent 81001FF Attenuation Filter before starting the
power tests in the following measurement setups; it is not sufficient to
use calibration factors that are derived from an earlier setup.
1 Make sure all instruments have warmed up.
Figure 18 Calibration of the Agilent 81001FF Attenuation Filter, Reference Setup
2 Set the power meter installed in the HP 8153A to the following
values:
a Set range to 0 dBm; press Up or Down as required.
b Set T, the averaging time, to 500 ms.
c Agilent 81680A, 40A, 82A, 42A, & 89A:
Set λ, the wavelength, to 1550 nm.
Agilent 81480A:
Set λ, the wavelength, to 1450 nm.
d Select dBm as the power units.
3 Move to the Tunable Laser channel of the Agilent 8164A.
Agilent 81680A, 40A, 82A, 42A, & 89A:
Set λ, the wavelength, to 1550 nm, and [P], the power, to 0 dBm..
For Agilent 81480A:
Set λ, the wavelength, to 1450 nm, and [P], the power, to 0 dBm.
4 Press the key beside the laser output to switch the laser on.
Agilent 8164A Lightwave
Measurement System
HP 81524A
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted
Connector
Optical Head
Agilent 81000SA
HP 81533B
HP 8153A Lightwave Multimeter