Agilent Technologies 81689A Stud Sensor User Manual


 
Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 69
Maximum Output Power
Make sure the instruments have warmed up before starting the
measurement.
NOTE Absolute Power Accuracy is not specified.
The result of the measurement below is greatly influenced by the
quality and the matching of the used interconnections.
1 Set up the equipment as shown in Table Figure 20.
Figure 20 Test Setup for the Maximum Output Power Tests
2 Set the Power Meter to the following settings:
a Select automatic ranging; press Auto as required.
b Set T, the averaging time, to 500 ms.
c Select dBm as the power units.
3 Move to the Tunable Laser channel of the Agilent 8164A Lightwave
Measurement System and press [Menu].
4 Set the menu parameters to the values shown in Table Table 5.
5 If you are using the Agilent 81480A, Agilent 81680A, or the
Agilent 81640A Tunable Laser module:
Connect the output fiber to Output 1, the Low SSE output,
remember to calibrate the Agilent 81001FF Attenuation Filter.
Set <Optical Output> to <Low SSE (1)>.
Agilent 8164A Lightwave
Measurement System
HP 81525A
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted
Connector
Optical Head
81000SA
HP 81533B
HP 8153A Lightwave Multimeter