Performance Test Instructions Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 73
Example (Agilent 81680A Output 1)
Power LinearityOutput 1
Power Linearity - High Power Test
Follow the steps below to measure the power linearity (without using
attenuation) of any one of the following:
Output2, the High Power output, of the Agilent 81480A
Output 2, the High Power output, of the Agilent 81680A
Output 2, the High Power output, of the Agilent 81640A
Agilent 81682A standard
Agilent 81682A #003
Agilent 81642A standard
Agilent 81642A #003
Agilent 81689A
1 Set up the equipment as shown in Figure 20 (remember to calibrate
the Agilent 81001FF Attenuation Filter if using 81524A).
2 Move to the Tunable Laser channel of the Agilent 8164A Lightwave
Measurement System and press [Menu].
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity
at current setting
Start = REF 6.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
7.0 dBm 1.02 dB + 1.00 dB = 0.02 dB
8.0 dBm 1.92 dB + 2.00 dB = +0.08 dB
9.0 dBm 3.02 dB + 3.00 dB = 0.02 dB
10.0 dBm 3.95 dB + 4.00 dB = +0.05 dB
11.0 dBm 5.07 dB + 5.00 dB = 0.07 dB
12.0 dBm 5.96 dB + 6.00 dB = +0.04 dB
13.0 dBm 7.05 dB + 7.00 dB = 0.05 dB
Maximum Power Linearity at current setting: +0.08 dB
Minimum Power Linearity at current setting: 0.07 dB
Total Power Linearity:
(Max Power Linearity Min Power Linearity) 0.15 dBpp