Performance Tests Performance Test Instructions
74 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
3 Set the menu parameters to the values shown in Table Table 5.
For Agilent 81480A, Agilent 81680A, Agilent 81640A,
Agilent 81682A, and Agilent 81642A tunable lase modules:
Set <Power Mode> to <Manual Att>.
4 Set the wavelength and power for each Tunable Laser module to the
values given in Table Table 12.
NOTE If you use the Agilent 81480A Output 2, or Agilent 81680A Output 2,
without attenuation, refer to the table “Power Linearity Output 2,
High Power Upper Power Levels” on page 120.
If you use the Agilent 81640A Output 2 without attenuation, use the
table “Power Linearity Output 2, High Power Upper Power Levels”
on page 134.
If you use the Agilent 81682A #003 without attenuation, use the
table “Power Linearity 81682A #003 Upper Power Levels” on
page 148.
If you use the Agilent 81642A #003 without attenuation, use the
table “Power Linearity 81642A #003 Upper Power Levels” on
page 160.
5 Perform the steps 5 to 16 of the “Power Linearity - Low Power Test”
on page 71.
Table 12 Wavelength and Power Settings for High Power Linearity Tests without
Attenuation
Module Wavelength [λ] Power [P] Attenuation [Atten]
Agilent 81480A - Output 2 1430.000 nm +3.000 dBm 0.000 dB
Agilent 81680A - Output 2 1540.000 nm +5.000 dBm 0.000 dB
Agilent 81640A - Output 2 1560.000 nm +2.000 dBm 0.000 dB
Agilent 81682A 1540.000 nm +6.000 dBm Not applicable
Agilent 81682A #003 1540.000 nm +4.500d Bm 0.000 dB
Agilent 81642A 1580.000 nm +6.000 dBm Not applicable
Agilent 81642A #003 1580.000 nm +4.500 dBm 0.000 dB
Agilent 81689A 1540.000 nm +6.000 dBm Not applicable