324 Chapter5
One-Button Measurement Functions
MEASURE (Spectrum Analysis Mode)
One-Button Measurement Functions
Remote Command:
Measurement Results Available
Command /
Condition
n Results Returned
:CONFigure:SEMask
:INITiate:SEMask
N/A Not Applicable
:FETCh:SEMask[n]?
:MEASure:SEMask[n]?
:READ:SEMask[n]?
Using Total Power
Reference
n=1 (or not
specified)
Returns 60 scalar results, in the following order:
1. Reserved for the future use, returns –999.0
2. Absolute power at the center frequency
(reference) area (dBm)
3. Reserved for the future use, returns –999.0
4. Reserved for the future use, returns –999.0
5. Peak frequency in the center frequency
(reference) area (Hz)
6. Reserved for the future use, returns –999.0
7. Reserved for the future use, returns –999.0
8. Reserved for the future use, returns –999.0
9. Reserved for the future use, returns –999.0
10.Reserved for the future use, returns -999.0.
11.Relative power on the negative offset A (dBc)
12.Absolute power on the negative offset A (dBm)
13.Relative peak power on the negative offset A
(dBc)
14.Absolute peak power on the negative offset A
(dBm)
15.Peak frequency in the negative offset A (Hz)
16.Relative power on the positive offset A (dBc)
17.Absolute power on the positive offset A (dBm)
18.Relative peak power on the positive offset A
(dBc)
19.Absolute peak power on the positive offset A
(dBm)
20.Peak frequency in the positive offset A (Hz)
21.Relative power on the negative offset B (dBc)
…
59.Absolute peak power on the positive offset E
(dBm)
60.Peak frequency in the positive offset E (Hz)