Alcatel Carrier Internetworking Solutions ACT 250 Paint Sprayer User Manual


 
Commissioning
Alcatel Vacuum Technology France - ACT 250 addendum
21
January 2001
Detailed description of RS commands
SBY
Syntax
Result
See also: NSP, RPM
Switches the speed set point to the stand-by value
#adrSBY<CR>[<LF>]
#adr,ok
Resets the stand-by speed to its last stored value, and allows it to be
modied if an RPM command is sent.
This conguration is automatically stored in user memory.
SEL
Syntax
Result
Returns the state of the parameters dened by OPT
#adrSEL<CR>[<LF>]
#adr,a,u
a: Reserved
u: Returns the choice of temperature unit::
u = 0: degrees Centigrade u = 1: degrees Fahrenheit
Denes the character which separates the parameters
in a reply
#adrSEPnnn<CR>[<LF>]
3-digit decimal value of the ASCII code of the desired char-
acter (with leading zeros if necessary).
condition: 000 nnn 255
#adr,ok or #adr,ErrX if error
Allows the user to select the character which separates
the parameters returned by the DLR, STA and LEV
commands. Default value: comma , ASCII code = 044
If ok, the selected value is automatically stored in user memory.
SEP
Syntax
Result