Emerson 403VP Stud Sensor User Manual


 
Instruction Manual 403 and 403VP
LIQ-MAN-ABR-403-403VP February 2015
PURSense
Conductivity Sensors
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process com position and operating conditions.
Application compat ibility is entirely the responsibility of the user.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process
temperature is lowered to a safe level!
Specifications - Sensor
Specifications 403 and 403VP
Wetted Materials titanium, PCTFE, 316 SS, EP
Temperature Range
32–221 °F (0–105 °C) Sensors tolerate
steam sterilization to 135 °C
Maximum Pressure 250 psig (1825 kPa abs)
NOTE
Elastomers and uorocarbon resins are compatible with 21CFR177. Elastomers also meet the requirements of USP
Class VI. Stainless steel contains <5 % delta ferrite. All surfaces in have 16 microinch (0.4 micrometer) Ra nish.
Installation
Depending on the option selected, the sensor can be installed in either a 1 ½-inch or 2-inch Tri-Clamp tee. The gasket,
clamp, and tee must be supplied by the user. The electrodes must be completely submerged in the process liquid, i.e., up
to the inside surface of the ange.
If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the
electrodes. Trapped bubbles will cause errors. Normally, as bubbles accumulate the conductivity reading drifts down. To
control bubble formation, apply a small amount of back pressure to the sensor.
For additional information, please visit our website at www.rosemountanalytical.com.
Figure 1. Sensor Orientation
°
°
Figure 2. Recommended Installation