Fluke PM-3370B Sander User Manual


 
4 - 24 COMMAND REFERENCE
Cancels or aborts any instrument-dependent action.
Cancels the effect of the
*
OPC command and the
*
OPC? query.
Sets the TRIGger subsystem into its IDLE state.
The
*
RST command does not affect the following:
State of the IEEE 488.1 interface.
GPIB (IEEE 488.1) address of the instrument.
Contents of the Output Queue.
Contents of the Error/Event Queue.
Service Request Enable setting in the SRE register.
Transition filters in the status subsystem.
Event registers in the status subsystem.
Event enable registers in the status subsystem.
Calibration data that affects the device specifications.
Version number set by the SYSTem:VERSion command.
Contents of the internal memory registers (
*
SAV/
*
RCL).
Example:
Send
*RST
Front panel compliance:
All settings not mentioned in the description are set according to the front panel
fixed setup, which can be recalled by pressing the keys STATUS and TEXT OFF
at the same time.
FUNCTION: DEFAULT SETTING(S):
TB mode
Acquire
Acquisition
Pre-trigger view
Bandwidth limiter
Measure 1 & 2
Math 1 & 2
Cursors
Trace intensity
User text
Display
Beeper
Hardcopy PRINT & PLOT
Pass/Fail testing
Realtime only OFF
Event delay OFF
Acquisition length 512 (samples of 16 bits)
Trigger Level MAX
Averaging OFF
Peak detection OFF
Envelope OFF
Autoranging attenuators OFF
Locked
50% of MTB (-5 ms)
OFF
OFF
OFF
OFF
0.18
Data cleared
OFF
ON
Plotter; HPGL
OFF