Fluke PM-3370B Sander User Manual


 
4 - 78 COMMAND REFERENCE
SENSe:FUNCtion:OFF
SENSe:FUNCtion[:ON]
SENSe:FUNCtion:STATe?
Syntax:
SENSe:FUNCtion:OFF "XTIMe:VOLTage<n>"
SENSe:FUNCtion:OFF "XTIMe:VOLTage:SUM <i,j>"
SENSe:FUNCtion[:ON] "XTIMe:VOLTage<n>"
SENSe:FUNCtion[:ON] "XTIMe:VOLTage:SUM <i,j>"
<n> [1] | 2 | 3 | 4
1 = CH1, 2 = CH2, 3 = CH3, 4 = CH4
Note: CH3 not applicable for PM33x0B.
<i,j> 1,2 | 3,4
1,2 CH1 + CH2
3,4 CH3 + CH4 (not for PM33x0B)
Query form:
SENSe:FUNCtion:STATe? "XTIMe:VOLTage<n>"
Response:
0 | 1
0 Input channel <n> is off.
1 Input channel <n> is on.
Query form:
SENSe:FUNCtion:STATe? "XTIMe:VOLTage:SUM <i,j>"
Response:
0 | 1
0 Addition of channel i+j is off.
1 Addition of channel i+j is on.
Description:
The SENSe:FUNCtion[:ON] command switches the input channel specified by
<n> or the addition of two input channels specified by <i,j> ON.
The SENSe:FUNCtion:OFF command switches the input channel specified by
<n> or the addition of two input channels specified by <i,j> OFF.
The SENSe:FUNCtion:STATe? query reports whether the specified input channel
<n> or the addition of the input channels <i,j> is ON or OFF.