Fluke PM-3370B Sander User Manual


 
COMMAND REFERENCE 4 - 133
Description:
The TRIGger:VIDeo:LINE command selects the video line number. Depending on
the video system selected, the following ranges are valid:
> NTSC from 1 to 525
> PAL or SECAM from 1 to 625
> HDTV from 1 to 1250
The TRIGger:VIDeo:SSIGnal command selects the video signal polarity.
After a
*
RST command, video line number 1 and signal polarity POSitive are
selected.
Example:
Send TRIGger:TYPE VIDeo Selects TV video triggering.
Send
TRIGger:VIDeo:LINE 123 Selects video line number
123.
Send
TRIGger:VIDeo:SSIGnal NEGative Selects negative video
signal polarity.
Front panel compliance:
The TRIGger:VIDeo:LINE command is the remote equivalent of the front panel
TRIGGER MAIN TB - LINE NBR softkey menu. The TRIGger:VIDeo:SSIGnal
command is the remote equivalent of the front panel TRIGGER MAIN TB -
pos/neg softkey menu.