Maintenance
6–8
THS710A, THS720A, THS730A & THS720P Service Manual
You will need a
1
@
4
inch flat-bladed screwdriver to remove the tilt stand.
STOP. You do not have to remove the tilt stand to access internal modules. In
most cases, you do not have to remove the tilt stand to clean the TekScope
instrument exterior.
Removal. Use this procedure to remove a defective tilt stand.
1. Place the TekScope instrument face down on a soft surface (such as an
anti-static mat).
2. Insert the
1
@
4
inch flat-bladed screwdriver into the gap between the tilt stand
and instrument case as shown in Figure 6–1.
3. Twist the screwdriver gently to release one side of the tilt stand from the
instrument case. Take care not to damage the case.
Figure 6–1: Removing the tilt stand
Tilt Stand