Command Errors (continued)
141 Invalid character data
Either the character data element contains an invalid character, or the element is not valid.
144 Character data too long
The character data element contains more than 12 characters.
148 Character data not allowed
A discrete parameter was received, but a string or numeric parameter was expected.
150 String data error
Generic string data error
151 Invalid string data
An invalid character string was received. Check that the string is enclosed in quotation marks.
158 String data not allowed
A character string was received, but is not allowed for this command.
160 Block data error
Generic block data error
161 Invalid block data
The number of data bytes sent does not match the number of bytes specified in the header.
168 Block data not allowed
Data was sent in arbitrary block format but is not allowed for this command.
170 Expression error
Generic expression error
171 Invalid expression data
The expression data element was invalid.
178 Expression data not allowed
Expression data element was sent but is not allowed for this command.
Execution Errors (these errors set Standard Event Status register bit #4)
200 Execution error
Generic syntax error
220 Parameter error
A data element related error occurred.
221 Settings conflict
A data element could not be executed because of the present instrument state.
222 Data out of range
A data element could not be executed because the value was outside the valid range.
223 Too much data
A data element was received that contains more data than the instrument can handle.
224 Illegal parameter value
An exact value was expected but not received.
225 Out of memory
The device has insufficient memory to perform the requested operation.
226 Lists not same length
One or more lists are not the same length.
230 Data corrupt or stale
Possible invalid data. A new reading was started but not completed.