E
Manual Backdating
Introduction
The backdating information in this chapter applies to units that have the following serial numbers:
Agilent Model 6625A. serials 2831A-00101 to 00663
Agilent Model 6626A, serials 2831A-00101 to 00823
Agilent Model 6628A, serials 2922A-00101 to 00232
Agilent Model 6629A, serials 2920A-00101 to 00292
Make Changes
On page 29, replace the information in Line Voltage Conversion paragraph under steps number 2, 3, and 4 as follows:
2. To open the line module push against the tab on the line module in the direction of the ac input socket.
3. Remove the voltage selector card from its receptacle. This card is about one inch square and has a red notch in
each of its four edges.
4. To select a voltage, insert the card into its receptacle so that after the door is closed, the red notch shows through
the window next to the voltage level you require.
It is possible to insert the voltage selector card so that a red notch is not visible through any of the four
windows. Do NOT operate the power supply while the selector card is in this position.
Also on page 28, replace figure 2-3 with the following figure;
Figure 2-3. Line Module Detail