Invacare IRC 1705 Air Compressor User Manual


 
SUPPLEMENTAL INFORMATION
Invacare
®
Select 14 Part No 1148073
SUPPLEMENTAL
INFORMATION
Compressor
If the unit shuts down frequently, you may have a low voltage
situation.
Low voltage can also be suspected when:
The motor does not get up to full power or speed.
Home fuses or circuit breakers activate when starting
compressor.
Lights dim or remain dim when compressor is started.
Other motor operated appliances fail to operate properly
or too many motor operated appliances are on same
circuit.
Extension Cords
If you must use an extension cord, use only a cord having the
same or higher electrical rating as the device being connected.
Make certain the extension cord is in good condition without
any exposed wires or hazardous contacts, and the extension
cord wire is NOT LESS than 18 gauge for 25 feet.