Omega FCLTX-100 Stud Sensor User Manual


 
PRODUCT INSTRUCTION SHEET
Operating temperature range: 0-45 degC
Maximum operating pressure: 1 bar/14.7 psi/1atm
Flow rate minimum: .20 gpm
Flow rate maximum: 0.6 ppm
pH range: 5.5-9.0
Output signal: 4.0+/- 0.2mA in air (zero)
20mA +/- 0.2mA at high range
(2, 5 or 10ppm)
Power Requirement: 12-24 VDC, 250 mA minimum
Cross-Sensitivity: ClO2, ozone, bromine
Chemical Compatibility: up to 50% ethanol/water or
up to 50% glycerol/water
OPERATING SPECIFICATIONS
SECTION 12.0
SENSOR SPECIFICATION
12.1 OPERATING SPECIFICATIONS. Follow all operating specifications,
especially for pH and flow rate as noted in the specification tables
below.
SENSOR
Dimensions: 8.2"L x 1" dia
Body Material: Black PVC
Membrane Material: polyethersulfone
O-ring material: Viton
R
Cathode: Gold
Anode: Silver chloride (AgCl)
Cable: 2 -conductor shielded, 10ft (3mtr)
tinned wire leads
FLOW CELL
Dimensions: 5.58"H x 2.25"DIA
Material: Acrylic
Connections: 1/4" NPT inlet and outlet
TECHNICAL SPECIFICATIONS
MAINTENANCE/REPLACEMENT PARTS
FC72 Flow Cell, 1/4 inch FNPT inlet and outlet, includes:
2 each 1/4" barbed tube fittings, clamp, threaded
flow cell installation fitting
FCL-ES Free Chlorine sensor fill solution, 30mL, 1 each
FCL-RM FCL Replacement premembraned cap, 1 each
FC72 FLOW CELL DIMENSIONS
SENSOR AND FLOW CELL INSTALLATION DIMENSIONS
Parts covered by this product instruction sheet include: FCLTX-100 Series