Omega FTB600B Stud Sensor User Manual


 
10
5
FTB600B Series
Installation
Section 5 - Installing the FTB600B Series in the fluid line.
A. Make sure the fluid is compatible with PVDF (polyvinylidene fluoride)
and meets viscosity, pressure and temperature parameters of the
FTB600B Series. The fluid must also meet filtration requirements as
listed in 3D.
B. Install the FTB600B series in the fluid line with the arrow pointing in
the direction of the flow.
C. While installing the FTB600B Series in the fluid line be careful not to
over-torque the end caps (on hose-barbed flowmeters) or other
fittings on the flowmeter. Due to the relatively soft composition of
PVDF the body or threads can be permanently distorted.
D. Attach wires to the readout display with the display and power off.
Not only will this help to avoid a potential shock hazard, but it can
also help prevent an error in hooking the flow meter to an incorrect
115 Vac supply.
E. Connect digital display to power supply and enter scaling factors for
both the rate and total. Follow the manufacturers instructions for
programming the digital display.
F. The FTB600B Series is now ready for use.
G. See Table 5-1 for straight pipe recommendations.