Omega OS35 Stud Sensor User Manual


 
OS35
Infrared Temperature Sensor
Installation Instructions
1. Install OS35 as close as practical to view target material to be measured
2. Wire OS35 with Red to Positive Power, Black to Ground, White to Positive
Output Signal, and Green to Reference Output Signal
3. Measure actual temperature of target material with OS950 Thermometer or
any reliable reference
4. Adjust “offset”, “zero”, or “low cal” on input device to match reading to ac-
tual thermometer reading
5. Installation calibration complete
*Six counts off one count on
**Six counts on one count off
ERROR MESSAGES
Condition Priority LED Display 0-5V 0-10V 4-20mA RS-232
Low Power 1 OFF Under 0.1V Under 0.1V Under 4mA Not Implemented
Hardware Internal
Errors
2, 13 Uniform Flash Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Vsig-Offset High 3 Uniform Flash Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Vsig-Offset Low 4 Uniform Flash Over 4.9V Over 9.8V Over 19.7mA Not Implemented
EMI 5 Uniform Flash Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Range Error 6 Uniform Flash Over 4.9V Over 9.8V Over 19.7mA Not Implemented
High Ambient 7 Long Flash** Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Low Ambient 8 Short Flash* Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Too Much Heat
Flow
9 Long Flash** Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Too Little Heat
Flow
10 Short Flash* Over 4.9V Over 9.8V Over 19.7mA Not Implemented
High Target 11 Long Flash** Over 4.9V Over 9.8V Over 19.7mA Not Implemented
Low Target 12 Short Flash* Over 4.9V Over 9.8V Over 19.7mA Not Implemented