Philips Trilogy 200 Work Light User Manual


 
Trilogy200 clinical manual
84
Note: The Flow Pattern setting
might be limited to only Ramp
or Square based on the Tidal
Volume, Inspiratory Time, and
Breath Rate settings to maintain
the minimum and maximum
peak ows.
3. PEEP
The Positive End Expiratory Pressure (PEEP) setting can be
increased from 0 to 25 cm H
2
O in active circuits and 4 to 25 cm
H
2
O in passive circuits, in increments of 1. PEEP is the positive
pressure maintained in the patient circuit during exhalation. The
PEEP must be less than or equal to the pressure setting.
Control Ventilation (CV) Mode
The following settings, described in the previous sections of this
chapter, are also available in CV mode:
Tidal Volume
Breath Rate
Inspiratory Time
PEEP
High Inspiratory Pressure
Low Inspiratory Pressure
In addition to these, the following settings are also available in CV
mode.
1. Flow Pattern
You can choose either Ramp or Square for the Flow Pattern setting.
2. Sigh
You can enable or disable the Sigh setting by selecting On or O.
A Sigh is a breath that is delivered every 100 breaths at 150% of
the normal volume.