Tektronix TDS724C Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–91
CURSOR:VBARS SNAP
specifies that the cursors positions are the same as the current DATA:START and
DATA:STOP values.
CURSOR:VBARS?
might return :CURSOR:VBARS:UNITS SECONDS;POSITION1 1.00E-6;POSIĆ
TION2 9.00E-6;SELECT CURSOR2.
CURSor:VBArs:DELTa? (Query Only)
Returns the time or frequency between the two vertical bar cursors. The units,
seconds or Hertz, are specified by the CURSor:VBArs:UNIts command.
The position can appear in units of seconds, 1/seconds (hertz), or video line
numbers (with option 05). On the TDS 400A, position can also appear in units of
clocks or 1/clocks.
Cursor
CURSor:VBArs:UNIts
CURSor:VBArs:DELTa?
CURSor
:
VBArs
:
DELTa
?
<NR3>
CURSOR:VBARS:DELTa?
might return 1.064E+00, indicating that the time between the vertical bar cursors
is 1.064 seconds.
CURSor:VBArs:POSITION<x>
Positions a vertical bar cursor for both vertical bar and paired cursors. The
CURSor:VBArs:UNIts command specifies units.
The position can appear in units of seconds, 1/seconds (hertz), or video line
numbers (with option 05). On the TDS 400A, position can also appear in units of
clocks or 1/clocks.
Examples
Group
Related Commands
Syntax
Returns
Examples