![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/6/1a/61a2dc46-154d-4da2-92af-3f6888cccd18/61a2dc46-154d-4da2-92af-3f6888cccd18-bg47.png)
5-17
Servicing
Troubleshooting
To check the A2A2 Laser Modulation Board Assembly
1
Check the +5V supply by probing the center pin of J11. The modulation input
cable from the rear panel connects to this jack.
2
Check the –15V supply by probing any of the three resistors that are located
next to J2. These resistors are loaded in a straight line and have the label “1471”
printed on them.
3
Locate the large 3W black resistor on the assembly that is closest to the pump
laser.
4
Measure the voltage on the end of the resistor that is furthest from the pump
laser. The voltage should measure approximately –1V to –2V.
5
The pump laser is being driven by the current source and should be turned on.
6
Press the front-panel
ACTIVE
key so that the front-panel
ACTIVE
light is turned
off. The voltage measured at the 3W resistor should now be approximately
+0.7V.
The pump laser is turned off.
7
If the voltage measured across the resistor is incorrect, return the instrument
to Agilent Technologies for servicing.