A SERVICE OF

logo

5-17
Servicing
Troubleshooting
To check the A2A2 Laser Modulation Board Assembly
1
Check the +5V supply by probing the center pin of J11. The modulation input
cable from the rear panel connects to this jack.
2
Check the 15V supply by probing any of the three resistors that are located
next to J2. These resistors are loaded in a straight line and have the label 1471
printed on them.
3
Locate the large 3W black resistor on the assembly that is closest to the pump
laser.
4
Measure the voltage on the end of the resistor that is furthest from the pump
laser. The voltage should measure approximately 1V to 2V.
5
The pump laser is being driven by the current source and should be turned on.
6
Press the front-panel
ACTIVE
key so that the front-panel
ACTIVE
light is turned
off. The voltage measured at the 3W resistor should now be approximately
+0.7V.
The pump laser is turned off.
7
If the voltage measured across the resistor is incorrect, return the instrument
to Agilent Technologies for servicing.