SPECIFICATIONS
Ranges Resolution Accuracy
pH 0.00to14.00pH 0.01pH ±0.02pH
Conductivity
0.0to200.0µS
200to2000µS
2.00to20.00mS
0.1µS
1µS
0.01mS
±2%FS
TDS 0.0to134.0ppm
134to1340ppm
1.34to13.40ppt
0.1ppm
1ppm
0.01ppt
(calculatedfromConductivity)
Salinity 0.0to100.0ppm
100to1000ppm
1.00to10.00ppt
0.1ppm
1ppm
0.01ppt
(calculatedfromConductivity)
ORP‐1500to1500mV 1mV ±3mV
Temperature 32to194ºF
0.0to90.0ºC
0.1º≤99.9º
≥100º
±2ºF/1ºC(meter+probe)
Display 9999countLCD
MTCtemperaturerange 32.0to194.0ºF(0.0to90.0ºC)
pHcalibrationpoints 4.00,7.00and10.00pH
Conductivitycalibrationpoints 84.0µS,1413µS,12.88mS
TDSconversionratio 0.67fixed
Salinityconversionratio 0.5fixed
AutoPowerOFF 10minutes,withdisable
Overrangeindication “OL”
OperatingTemperature 41ºFto
104ºF(5ºCto40ºC)
StorageTemperature‐4ºFto140ºF(20ºCto60ºC)
OperatingHumidityMax80%upto87ºF(31ºC)decreasinglinearlyto50%at
104ºF(40ºC)
StorageHumidity <80%
OperatingAltitude 7000ft.(2000meters)maximum.
Power 9ValkalinebatteryorACadapter
Dimensions 4.7x3.8x1.8”(118x96x45mm)closed
Weight 12oz.(340g)