Specifications
Display 5 digit multifunction LCD
Output (source) Ranges, Resolution, and Accuracy
DC Current
4 to 20.000mA, 0 to 20.000mA, 0 to 24.000mA
Accuracy for all mA ranges: ± (0.025% + 3 digits)
DC Voltage
0 to 100.00mV; 0 to 1.0000V; 0 to 12.000V;
Accuracy for all voltage ranges: ± (0.05% + 3 digits)
Type K Temperature
-328 to -148
o
F (-200 to -100
o
C); Accuracy: ±3.6
o
F (2.0
o
C)
-148 to 32
o
F (-100 to 0
o
C); Accuracy: ±2.0
o
F (1.1
o
C)
32 to 2400
o
F (0 to 1370
o
C); Accuracy: ±1.5
o
F (0.8
o
C)
Type J Temperature
-148 to 32
o
F (-100 to 0
o
C); Accuracy: ±1.6
o
F (0.9
o
C)
32 to 1400
o
F (0 to 760
o
C); Accuracy: ±1.2
o
F (0.7
o
C)
Type E Temperature
-148 to 32
o
F (-100 to 0
o
C); Accuracy: ±1.6
o
F (0.9
o
C)
32 to 1292
o
F (0 to 700
o
C); Accuracy: ±1.2
o
F (0.7
o
C)
Type T Temperature
-328 to 32
o
F (-200 to 0
o
C); Accuracy: ±1.8
o
F (1.0
o
C)
32 to 752
o
F (0 to 400
o
C); Accuracy: ±1.5
o
F (0.8
o
C)
Frequency
1 to 125 Hz; Accuracy: ± 0.04Hz
126 Hz to 62,500 Hz; Accuracy: ± (0.01% + 0.04Hz)
Accuracy note:
All published accuracies are 'of reading'. Meter must be warmed up for
20 minutes before use for optimum accuracy.
Power supply 9V battery, external battery pack (6 x 1.5V 'AA'),
or optional AC adapter;
Power Consumption 60 to 180mA
Operating Conditions 32 to 122
o
F (0 to 50
o
C); 85% RH max.
Storage Conditions -4 to 140
o
F (-20 to 60
o
C); 85% RH max.
Dimensions 3.46 x 6.61 x 1.03" (88 x 168 x 26mm)
Weight 11.63 oz. (330g)