10 SDL800-en-GB_V1.4 6/15
Specifications
Display Backlit LCD; LCD size: 52 x 38mm (2 x 1.5”)
Status indicators Over-range audible beep and low battery display icon
Frequency Range 10Hz to 1kHz
Datalogger Sampling Rate AUTO LOGGING: From 1 to 3600 seconds.
MANUAL LOGGING: Set the sampling rate to ‘0’ seconds
Memory Card SD memory card; 1G to 16GB size
Data Hold Freeze the displayed reading
Memory Recall Record and Recall the Maximum and Minimum readings
Display update rate Approx. 1 second
Data Output RS-232 / USB PC computer interface
Operating Temperature 0 to 50
°
C (32 to 122
°
F)
Operating Humidity 85% R.H. max.
Auto Power OFF After 10 minutes of inactivity (can be disabled)
Power Supply Six (6) 1.5 VDC batteries (optional 9V AC adaptor)
Power Consumption Normal operation (backlight and datalogger OFF): approx. 5mAdc
With backlight OFF and datalogging ON: approx. 25mAdc
With backlight ON add approx. 12mAdc
Weight 343g (0.75 lbs.) w/ batteries; 241g (0.53 lbs.) w/o batteries
Dimensions Meter: 182 x 73 x 47.5mm (7.1 x 2.9 x 1.9”)
Vibration sensor: round 16mm (0.63”) diameter, 37mm (1.46”)
length
Electrical Specifications
Function
Unit
Range
Accuracy
Acceleration
m/s
2
0.5 to 199.9 m/s
2
± ( 5%rdg + 2 d )
@ 80 and 160Hz
g
0.05 to 20.39 G
ft/s
2
2 to 656 ft/s
2
Calibration Point: 50 m/S^2 ( 160 Hz )
Velocity
mm/s 0.5 to 199.9 mm/s
± ( 5%rdg + 2 d )
@ 80 and 160Hz
cm/s 0.05 to 19.99 cm/s
inch/s 0.02 to 7.87 inch/s
Calibration Point: 50 mm/s ( 160 Hz )
Displacement
mm 0.003 to 1.999 mm
± ( 5%rdg + 2 d )
@ 80 and 160Hz
inch 0.078 inch
Calibration Point: 0.141 mm ( 160 Hz )
Above specification tests under the environment RF Field Strength less than 3 V/M & frequency less
than 30 MHz only and 23±0.5°
C
Copyright©2013‐2015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpart
inanyform.
ISO‐9001Certified
www.extech.com