LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03
21
TROUBLESHOOTING
Warning: If you experience a problem with your lift and are unable to service it yourself,
contact your Lumex dealer.
LF1030 HYDRAULIC PATIENT LIFT
SYMPTOM POSSIBLE CAUSE ACTION
Patient lift does not Air in the cylinder Bleed the cylinder by turning the release knob
go up or down and push down on the mast. Pump up the lift
Loss of hydraulic fluid again and turn the release knob again.
(leaking cylinder)
Replace any leaking cylinder
Patient lift does not Air in the cylinder See above Directions
go down
Note: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex dealer for a
new cylinder.
LF1040 ELECTRIC PATIENT LIFT
SYMPTOM POSSIBLE CAUSE ACTION
Control box is not Battery discharged Charge battery
responding and relay Battery charger defective Replace battery charger
is not clicking Battery defective Replace battery
Hand control defective Replace hand control
Actuator does not run, Actuator plug not properly Push actuator plug firmly into control box
but relay is clicking seated
Actuator defective Replace actuator
Control box defective Replace control box
Actuator goes up or Hand control defective Replace hand control
down, but not both Control box defective Replace control box