ERROR CODES
ERROR 0 - NO ERROR
ERROR 1- INVALID LOW-SCALE ADJUSTMENT OF RECORDER OUTPUT
1. The manual (fine) low-scale adjustment of the recorder
output in the DOWN direction (DIAG 101) could not be
performed.
2. The automatic low-scale adjustment of the recorder out-
put in the DOWN direction (DIAG 104 or 106) could
not be performed.
Check the dip switches for proper configuration.
Also, when in DIAG 101, “E01” can occur if the user does
not press DIAG again before the displayed DVM reading
goes below:
a. 100 mV, with 0-1 V dip switch setting
b. 10 mV, with 0-100 mV dip switch setting
c. 1 mV, with 0-10 mV dip switch setting.
Another possibility is hardware failure.
ERROR 2 - INVALID FULL-SCALE ADJUSTMENT OF RECORDER OUTPUT
1. The manual (fine) full-scale adjustment of the recorder
output in the UP direction (DIAG 102) could not be per-
formed.
2. The automatic full-scale adjustment of the recorder out-
put in the DOWN direction (DIAG 105 or 107) could
not be performed.
Check the dip switches for proper configuration.
Also, when in DIAG 102, “E02” can occur if the user does
not press DIAG again before the displayed DVM reading
goes above:
a. 1.1 V, with 0-1 V dip switch setting
b. 110 mV, with 0-100 mV dip switch setting
c. 11 mV, with 0-10 mV dip switch setting.
Another possibility is hardware failure.
ERROR 3 - INVALID LOW-SCALE ADJUSTMENT OF RECORDER OUTPUT
The manual (fine) low-scale adjustment of the recorder
output in the UP direction (DIAG 100) could not be per-
formed.