![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/b/49/b4926d49-60bb-4063-9668-a81619710d0f/b4926d49-60bb-4063-9668-a81619710d0f-bg80.png)
Nominal Traits
2–12
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Table 2–5: Nominal traits — Triggering system (cont.)
Name Description
n
, S
up
nd
d f
r
Feature Min to max
T
S
up
d V
n Tr
r
Setup Time
Hold Time
Setup + Hold Time
–100 ns to 100 ns
–1 ns to 100 ns
2 ns
For Setup Time, positive numbers mean a data transition before the clock edge and
negative means a transition after the clock edge.
For Hold Time, positive numbers mean a data transition after the clock edge and negative
means a transition before the clock edge.
Setup + Hold Time is the algebraic sum of the Setup Time and the Hold Time
programmed by the user.
n
, Tr
r
v
r T
r
d Source Range
Any Channel
Auxiliary
Line
±12 divisions from center of screen
±8 V
±400 V
Video Trigger Modes of Operation
(Option 05 Video Trigger)
Supports the following video standards:
H NTSC (525/60) – 2 field mono or 4 field
H PAL (625/50) – 2 field mono or SECAM, 8 field
H HDTV –
(787.5/60)
(1050/60)
(1125/60)
(1250/60)
H FlexFormat
TM
(user definable standards)
User can specify: field rate, number of lines, sync pulse width and polarity, line rate,
and vertical interval timing.