Agilent Technologies U2761A Welding System User Manual


 
82 U2761A User’s Guide
2Features and Functions
Remote Interface Operation
FREQuency:STARt {<frequency>}
FREQuency:STOP {<frequency>}
Sweep Mode
You can sweep with either linear or logarithmic spacing. For
a linear sweep, the U2761A varies the output frequency in a
linear fashion during the sweep. For a logarithmic sweep,
the U2761A varies the output frequency in a logarithmic
fashion.
The sweep mode consists of Linear or Logarithmic. The
default sweep mode is Linear.
Soft Front Panel Operation
On the Sweep Type panel as shown in Figure 2-28, select
the Linear or Log mode.
Remote Interface Operation
SWEep:SPACing {LINear|LOGarithmic}
Sweep Time
The sweep time specifies the number of seconds required to
sweep from the start frequency to the stop frequency. The
number of discrete frequency points in the sweep is
automatically calculated by the U2761A and is based on the
sweep time you have selected. The range of the sweep time
is 1 ms to 500 s. The default sweep time is 1 s.