Emerson LIQ_MAN_ABR_226 Stud Sensor User Manual


 
Instruction Manual 226
LIQ_MAN_ABR_226 April 2014
Submersion/Insertion Toroidal Sensor
For additional information, please visit our website at www.rosemountanalytical.com
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the responsibility of the user.
CAUTION
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process
temperature is lowered to a safe level!
Specifications
Wetted Material: PEEK (glass-lled), EPDM gasket
Temperature: 248 °F (120 °C) maximum
Pressure: 295 psig (2135 kPa abs)
Installation
Keep at least 2.4 in. (60 mm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor in place.
Ensure that the sensor is completely submerged in the liquid. Mounting the sensor in a vertical pipe run with the ow
from bottom to top is best. If the sensor must be installed in a horizontal pipe run, mount the sensor in the 3 o’clock or 9
o’clock position.
Figure 1. Submersion Installation Showing Mounting Hardware
Screw the rear facing thread of the PEEK adapter into the pipe. Do not let the back end of the sensor get wet.