9
Instruction Manual 226
LIQ_MAN_ABR_226 April 2014
Figure 13. Resistance check. Disconnect leads from analyzer before measuring resistances.
green
black
white
black
green - in
clear - common
white - sense
Resistance between shield and any other wire: >40 MΩ
1-2 Ω
1-2 Ω
1-2 Ω
-110 Ω
RTD
receive
226-54
drive
green
black
clear
white
black
clear
green - in
black - common
white - sense
clear - shield
drain
receive
RTD
1-2 Ω
1-2 Ω
1-2 Ω
-110 Ω
Resistance between shield and any other wire: >40 MΩ
226-56
Table 1.
Temperature Resistance
10 °C 103.9 Ω
20 °C 107.8 Ω
25 °C 109.7 Ω
30 °C 111.7 Ω
40 °C 115.5 Ω
50 °C 119.4 Ω