Fluke PM-3384B Sander User Manual


 
4 - 126 COMMAND REFERENCE
TRIGger[:SEQuence[1]]:TYPE
TRIGger[:STARt]:TYPE
Syntax:
TRIGger[:SEQuence[1]]:TYPE EDGE | VIDeo | LOGic
Alias:
TRIGger[:STARt]:TYPE EDGE | VIDeo | LOGic | GLITch
EDGE Selects edge triggering.
VIDeo Selects TV video triggering.
LOGic Selects logic triggering (only for
PM3384B-94B).
GLITch Selects glitch triggering (only for PM33x0B).
Query form:
TRIGger[:SEQuence[1]]:TYPE?
TRIGger[:STARt]:TYPE?
Response:
EDGE | VID | LOG
Description:
The TRIGger:TYPE command controls the type of triggering.
After a
*
RST command, the trigger type is EDGE (normal triggering).
Example:
Send
TRIGger:TYPE VIDeo Selects TV video triggering.
Front panel compliance:
The TRIGger:TYPE command is the remote equivalent of the front panel
TRIGGER MAIN TB -edge/tv/logic softkey menu.