Invacare IRC1175 Air Compressor User Manual


 
19
MAINTENANCE
Your Freeway Freedom, battery pack and power pack can be cleaned
by wiping with a damp cloth.
The carrying case can be machine washed (40
o
C). Tumble dry at low
temperature settings.
CLEANING THE SIDESTREAM NEBULIZER
NOTE: Refer to AFTER THE TREATMENT in the OPERATING section of this
manual for regular use cleaning instructions.
Initially and once each week your Sidestream nebulizer should be
boiled in water with two (2) drops of dish washing liquid for 5 to 10
minutes.
REPLACING THE AIR FILTER (FIGURE 4)
The air filter cleans the air that you use during treatment. The air filter
gets dirty over time and will need to be replaced. As soon as the air
filter no longer looks white and clean (approximately 3 months) it
should be changed.
NOTE: The air filter will need to be changed more regularly if used in an
environment with excessive dirt or smoke.
1. Remove the air filter cover and the dirty air filter.
2. Insert a clean air filter and replace the air filter cover.
3. Dispose of dirty air filter.
FIGURE 4 - REPLACING THE AIR FILTER
Air Filter Cover
Air Filter