8
NOTE: Monitor subsequent films on the same steps
selected for normal contrast index.
Date - Month - Day – Year
Processor Number - Processor identification
Emulsion Number - Film batch identification
Developer Type - Developer vendor
identification
Fixer Type - Fixer vendor identification
Film Type - Film vendor identification
Exposure Color - Exposure light (blue or green)
Developer Replenisher Rate - The rate of
developer replenishment
Fixer Replenisher Rate - The rate of fixer
replenishment
Processing Time - Film process time, input-to-
output
A box of film should be set aside from regular stock for
exclusive sensitometer use. New film stock will require
re-establishment of normal values because small density
changes are possible between film batches.
Daily Plotting of Data on Process Control
Record
Plot the Speed Index, Contrast Index, Base+Fog, and
Developer Temperature results on the process control
record each time a control film is developed. Record
data immediately so that it is not lost or changed.