Tektronix TDS684B Caulking Gun User Manual


 
Command Descriptions
2–316
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:SLEWRate:THReshold:LOW?
THReshold
TRIGger :
:
PULse
MAIn
:
:
?
<Space>
<NR3>
LOW
SLEWRate
<NR3> is the threshold, in volts.
TRIGGER:MAIN:PULSE:SLEWRATE:THRESHOLD:LOW 50E-3
sets the lower limit of the pulse slew rate trigger to 50 mV.
TRIGger:MAIn:PULse:SLEWRate:WHEn
TDS 500C, 600B, & 700C Only
Sets or queries whether to check for a slewing signal that is faster or slower than
the specified delta time. This is equivalent to selecting Trigger When from the
main Trigger’s Slew Rate menu and entering a value in the resulting side menu’s
Trig if Less Than or Trig if Greater Than items.
Trigger
TRIGger:MAIn:PULse:SLEWRate:WHEn { FASTERthan | SLOWERthan }
TRIGger:MAIn:PULse:SLEWRate:WHEn?
FASTERthan
TRIGger :
:
PULse
MAIn :
:
WHEN
SLEWRate
SLOWERthan
<Space>
?
FASTERthan sets the trigger to occur when the slew is faster than the set
volts/second rate.
SLOWERthan sets the trigger to occur when the slew is slower than the set
volts/second rate.
Arguments
Examples
Group
Syntax
Arguments