Tektronix TDS684B Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–73
CH<x>:IMPedance { FIFty | MEG }
CH<x>:IMPedance?
FIFty
IMPedance
<Space>
?
MEG
CH
:<X>
FIFty sets the specified channel to 50 W impedance.
MEG sets the specified channel to 1 MW impedance.
CH1:IMPEDANCE FIFty
establishes 50 W impedance on channel 1.
CH3:IMPEDANCE?
might return MEG, indicating that channel 3 is set to 1 MW impedance.
CH<x>:OFFSet
Sets or queries the offset, in volts, that is subtracted from the specified input
channel before it is acquired. The greater the offset, the lower on the display the
waveform appears. This is equivalent to setting Offset in the Vertical menu.
Vertical
CH<x>:POSition
CH<x>:OFFSet <NR3>
CH<x>:OFFSet?
CH
<x> :
OFFSet
?
<Space> <NR3>
<NR3> is the desired offset in volts. The range is dependent on the scale and the
probe attenuation factor. The offset ranges are shown below.
Syntax
Arguments
Examples
Group
Related Commands
Syntax
Arguments