50 Language Dictionary
Table 3-1. Link Parameter List
Parameter True Event Condition Valid for
CC Constant current event bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
CV Constant voltage event bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
ESB Standard event summary bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
LSC List sequence complete pulse
2
OUTP:TTLT:LINK TRIG[:STAR]:LINK
OPC Operation complete bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
OPER Operation summary bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
QUES Questionable summary bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
MAV Message available summary bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
OFF No linked event condition OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
RQS Request service summary bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
RTG Received a trigger bit
2
OUTP:TTLT:LINK TRIG[:STAR]:LINK
STC List step completed pulse
2
OUTP:TTLT:LINK TRIG[:STAR]:LINK
STS List step started pulse
2
OUTP:TTLT:LINK TRIG[:STAR]:LINK
SUM3 OPER or QUES or ESB bit
1
OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
TDC Trigger delay complete pulse
2
OUTP:TTLT:LINK TRIG[:STAR]:LINK
1
See “Chapter 4 - Status Reporting”
2
See “Chapter 5 - Synchronizing Power Module Output Changes”.
Table 3-2. Power Module Programming Parameters
Agilent Model
Parameter 66101A 66102A 66103A 66104A 66105A 66106A
Output Programming Range
(maximum programmable values):
Voltage:
8.190 V 20.475 V 35.831 V 61.425 V 122.85 V 204.75 V
Current:
16.380 A 7.678 A 4.607 A 2.559 A 1.280 A 0.768 A
OV Protection:
8.8 V 22.0 V 38.5 V 66.0 V 132.0 V 220.0 V
Average Resolution
Voltage:
2.4 mV 5.9 mV 10.4 mV 18.0 mV 36.0 mV 60.0 mV
Current:
4.6 mA 2.3 mA 1.4 mA 0.75 mA 0.38 mA 0.23 mA
OV Protection:
50 mV 120 mV 200 mV 375 mV 750 mV 1.25 V
*RST State Values
1
Voltage:
0
Current:
257 mA 120 mA 72 mA 40 mA 20 mA 12 mA
OV Protection:
8.8 V 22.0 V 38.5 V 66.0 V 132.0 V 220.0 V
1
These also are the power-on reset values when the factory default parameters are left in effect (see "Chapter 2
Installation” in the power module User’s Guide).
Nonvolatile Memory Locations:
5; (0 through 4)
Volatile Memory Locations:
5; (5 through 9)
: